In July 2022, our Calibration Laboratory received accreditation with the number AP 209 granted by the Polish Center for Accreditation (PCA) under the EA MLA international agreement. Thus, the Laboratory confirmed compliance with the requirements of PN-EN ISO/IEC 17025 "General requirements for the competence of testing and calibration laboratories" in the field of calibration of electrical quantities.
The Lutron PH206 is a digital PH meter in a rugged case. The device features a large LCD display.
Display | Dual-function, 13 mm (0.5") LCD display with adjustable contrast |
Measurement ranges | pH: 0 to 14 pH mV: 0 to 1999 mV temp.: 0 to 100 °C |
Input imp. | 1012W |
Temperature compensation for pH range | Manual: 0 to 100 °C, set byfront panel button; Automatic: with optional probe |
pH rangecalibration | Precision multi-turn SLOPE potentiometers (pH 4) and CAL. (pH 7) on the front panel |
pH probe | Any composite pH probe with BNC cable. |
Operating temp. | 0 to 50 °C |
Humidity | Maximum 80 % RH |
Sampling | every approx. 0.4 seconds |
Current consumption | approx. 2.7mA |
Weight and dimensions | 250g; 160 x 80 x 35 mm |