CHAUVIN ARNOUX C173 current probe
Parameters:
- Bandwidth: 10 Hz - 3 kHz
- Measurement of AC current from 1 mA to 1000 A in four ranges:
Range 1: 1 mA - 1 A (permissible measurement up to 120% of maximum nominal value)
Range 2: 0.01 A - 10 A (permissible measurement up to 120% of maximum nominal value)
Range 3: 0.1 A - 100 A (permissible measurement up to 120% of maximum nominal value)
Range 4: 1 A - 1000 A (permissible measurement up to 120% of maximum nominal value )
- Sensitivity: 1 V/A (range 1), 100 mV/A (range 2), 10 mV/A (range 3), 1 mV/A (range 4)
- Output signal:
Range1: 1 V AC per 1 A AC of measured current (1 V AC for 1 A AC)
Range2: 100 mV AC per 1 A AC of measured current (1 V AC for 10 A AC)
Range3: 10 mV AC per 1 A AC of measured current (1 V AC for 100 A AC)
Range 4: 1 mV AC per 1 A AC of measured current (1 V AC for 1000 A AC)
- Standard accuracy:≤0.7% to ≤3%(range 1); ≤0.5% to ≤1% ( range 2); ≤0.3% to ≤1% (range 3); ≤0.2% to ≤1%(range 4) - depending on the measured current value in the range
- IP40-rated housing
- Electrical safety: 600 V CAT III, 300 V CAT IV (IEC 1010-1 and IEC 1010-2-032)
- Maximum diameter of the measured wire: 52 mm
- Maximum dimension of the measured rail: 50 mm x 5 mm or 4 rails 30 mm x 5 mm
- Output: two-wire cable of approx. 1.5 m length terminated with two safe, male "banana" type terminals of 4 mm diameter
- Dimensions: 216 mm x 111 mm x 45 mm
- Weight: approx. 550 g